ST. JUDE MEDICAL, INC. INTEGRATED VIEWING SYSTEM, VANTAGEVIEW SYSTEM, (120VAC); DISPLAY, CATHODE-RAY TUBE, MEDICAL
|
Back to Search Results |
|
| Model Number |
VS100001 |
| Medical Device Problem Code |
Unexpected Shutdown (4019)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/13/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
During the procedure with the patient on the table, the system shutdown / went black and the case was cancelled.
|
| |
|
Additional Manufacturer Narrative
|
|
Additional information: g3, h2, h3 during field service, a faulty svdc power cable connection was replaced.Additionally, a faulty sc male fiber connector was noted on the blue fiber cable at the patch panel which was replaced.Replacement of the faulty power cable & connector resolved the issue and the system functioned as intended.This device history record was completed by a representative from avi.Based on the information provided to abbott and the investigation performed, the cause for the reported system shutdown was isolated to the power connector.
|
| |
|