H3, h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A visual inspection revealed the device was returned outside of original packaging.The inner blade was appears to have the silver plating flaked off around the distal tip of the device.No other visible damage to the device.A functional evaluation revealed the device functions as expected.The device made no grinding noise.The complaint was confirmed, the root cause was associated with a component failure.Factors that could have contributed to the reported event include an application of unintended inappropriate, excessive force, or wear of the device.No containment or corrective actions are recommended at this time.
|