Model Number 4002C0935
Medical Device Problem Code Difficult or Delayed Separation (4044)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/13/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The product was not returned for evaluation.Without the return of the device, the root cause of the problem cannot be determined.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.This report is associated with mfr report numbers: 3005168196-2021-01050, 3005168196-2021-01051.
 
Event or Problem Description
The patient was undergoing a coil embolization procedure in the splenic artery using penumbra coil 400s (pc400s), a penumbra coil 400 detachment handle (handle), and a non-penumbra microcatheter.During the procedure, while attempting to detach a pc400 using the handle, the coil alignment zone separated but the coil did not detach.It was reported that the click of the handle was not audible.The physician then made another attempt to detach the pc400 using the handle; however, the pusher assembly of the pc400 became stuck in the handle.Therefore, the physician manually detached the pc400.Next, the physician successfully detached another pc400 in the target vessel using another handle.Then, while attempting to advance the next pc400 through the microcatheter, the pc400 unintentionally detached and became stuck in the microcatheter.Therefore, the microcatheter containing the detached pc400 was removed from the patient.The procedure ended at this point.There was no report of an adverse effect to the patient.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PENUMBRA COIL 400
Common Device Name
HCG, KRD
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11828688
Report Number3005168196-2021-01052
Device Sequence Number1058678
Product Code HCG
UDI-Device Identifier00814548010588
UDI-Public00814548010588
Combination Product (Y/N)Y
Initial Reporter CountryIN
PMA/510(K) Number
K120330
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Reporter Type Manufacturer
Report Source distributor,foreign,health pr
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 04/19/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number4002C0935
Device Catalogue Number4002C0935
Device Lot NumberF101887
Was Device Available for Evaluation? No
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/19/2021
Initial Report FDA Received Date05/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age45 YR
Date Report Sent to FDA01/01/2005
-
-