SMITH & NEPHEW, INC. VULCAN GENERATOR CE MARK; ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES Back to Search Results
Model Number 7210812
Medical Device Problem Code Connection Problem (2900)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).
 
Event or Problem Description
It was reported that the vulcan generator did not recognize the return electrode.No case reported; therefore, there was no patient involvement.
 
Additional Manufacturer Narrative
The reported device, intended for use in treatment, was received for evaluation.There was no relationship found between the returned device and the reported incident.A visual inspection was performed on the product and no issue was observed.The complaint of return ground failure could not be confirmed.Product passed functional testing and all functions perform as expected.Faults or errors did not occur during functional testing.The complaint was not confirmed and the root cause could not be determined since the reported malfunction could not be duplicated during the product evaluation process.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.No containment or corrective actions are recommended at this time.
 
Additional Manufacturer Narrative
The reported device, intended for use in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VULCAN GENERATOR CE MARK
Common Device Name
ELECTROSURGICAL, CUTTING & COAGULATION & ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key11828712
Report Number1643264-2021-01857
Device Sequence Number13718261
Product Code GEI
UDI-Device Identifier03596010519399
UDI-Public03596010519399
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K991140
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 09/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number7210812
Device Catalogue Number7210812
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/05/2021
Initial Date Received by Manufacturer 04/26/2021
Supplement Date Received by Manufacturer06/16/2021
09/27/2021
Initial Report FDA Received Date05/14/2021
Supplement Report FDA Received Date06/17/2021
09/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-