• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. GRIPPER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. GRIPPER; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 71674080
Medical Device Problem Codes Break (1069); Fitting Problem (2183)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/29/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during trauma surgery, the inner workings of the gripper was broken and beyond repair.It was unable to grip onto guide rod.It was used vice grips instead (change in surgical technique).No injury o significant delay reported.
 
Additional Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.A complaint history review found related failures for the listed device; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the risk management file revealed this failure mode was previously identified.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GRIPPER
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11828848
Report Number1020279-2021-04344
Device Sequence Number3535267
Product Code LXH
UDI-Device Identifier03596010560766
UDI-Public03596010560766
Combination Product (Y/N)N
Initial Reporter StateIL
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 08/30/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number71674080
Device Catalogue Number71674080
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 04/29/2021
Supplement Date Received by Manufacturer08/26/2021
Initial Report FDA Received Date05/14/2021
Supplement Report FDA Received Date08/30/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-