| Model Number |
DIRECT BILIRUBIN, R1 4X6 ML, R2 4X6 ML |
| Medical Device Problem Code |
High Test Results (2457)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/28/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Customer consulted with chemistry systems support (css).Customer confirmed the erroneous results occurred with neonatal samples.Css were able to find out from the customer that the customer was using dbil reagent osr6111.Css recommended the customer change to osr61181, as osr61181 is for neonatal samples.The issue occurred due to use error, no further issues have been reported.Beckman coulter internal identifier is case-(b)(4).No patient demographic information (age, gender, weight, race or ethnicity) was provided.Customer phone number: (b)(6).
|
| |
|
Event or Problem Description
|
|
The customer reported receiving erroneous high patient results for direct bilirubin (dbil) on the au480 clinical chemistry analyzer for neonatal samples.The erroneous results were reported outside of the lab.There was a report of a change to patient treatment.Patients had a prolonged hospital stay with additional follow up testing/monitoring.It is unknown if additional treatment was received.
|
| |
|