• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC SOFAMOR DANEK USA, INC LONGITUDE II; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC SOFAMOR DANEK USA, INC LONGITUDE II; ORTHOPEDIC MANUAL SURGICAL INSTRUMENT Back to Search Results
Model Number 5584099
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Country: (b)(6) pma/510(k)#: class i exempt neither the device nor films of applicable imaging studies were returned to the manufacturer for evaluation.Therefore, we are unable to determine the definitive cause of the reported event.If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
Information was received from healthcare provider (hcp) via a manufacturer representative regarding a patient who was implanted with a spinal product type for lumbar spinal canal stenosis.It was reported that the tip of 4 drivers was broken when pedicle screw was inserted.There were no patient symptoms or complications as a result of this event.There were no fragments of the implants or instruments remained in the patient's body.There was a delay of less than 60min in overall procedure time.The screw placement was completed and the operation had been completed.
 
Additional Manufacturer Narrative
Additional information: d9, h3, h6 h3.Device evaluation summary: product analysis #(b)(4): part # 5584099 lot # nm15k018 visual inspection confirmed the entire torx tip of instrument has been sheared off.Optical examination of the fracture surface revealed a fairly flat fracture surface with circular material flow, consistent with torsional overload.This type of damage is consistent with torsional overload.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LONGITUDE II
Common Device Name
ORTHOPEDIC MANUAL SURGICAL INSTRUMENT
Manufacturer (Section D)
MEDTRONIC SOFAMOR DANEK USA, INC
4340 swinea rd
memphis TN 38118
MDR Report Key11830659
Report Number1030489-2021-00625
Device Sequence Number14400311
Product Code LXH
UDI-Device Identifier00643169163065
UDI-Public00643169163065
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 10/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number5584099
Device Catalogue Number5584099
Device Lot NumberNM15K018
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2021
Supplement Date Received by Manufacturer09/21/2021
Initial Report FDA Received Date05/17/2021
Supplement Report FDA Received Date10/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-