| Model Number |
CSE-P-125 |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
Unspecified Blood or Lymphatic problem (4434)
|
| Date of Event |
04/16/2021
|
|
Type of Reportable Event
|
Serious Injury
|
|
Additional Manufacturer Narrative
|
|
Customer informed haemonetics that they would not be able to provide further patient information.
|
| |
|
Event or Problem Description
|
|
On 21apr2021, haemonetics was notified of a long empty alert on cycle 5 of a vascular procedure in (b)(6), utilizing the cell saver® elite® autotransfusion system and cell saver® elite set - 125ml system.During procedure 2 units of donated blood was needed to preclude permanent impairment where it would not have been needed if the issue with the device did not occur.Although it was not able to be confirmed with the operator that there was an inner core leak this error has in the past been an indicator of a potential issue with the bowl that has been reported.Awaiting lot number.There is no additional information available at this time.
|
| |
|