SMITH & NEPHEW, INC. SEMIEXTENDED DRILL GUIDE; ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES Back to Search Results
Model Number 71654524
Device Problem Fitting Problem (2183)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/29/2021
Event Type  malfunction  
Event Description
It was reported that during trauma surgery the semiextended drill guide barely fits into the meta nail anterior drop.Surgeon was forced to hammer the drop into the guide.The drop is not the issue, because the standard drill guide fits perfectly fine into the anterior drop.The procedure was finished with the same device.No injury or significant delay reported.
 
Manufacturer Narrative
H3, h6: the device, used in treatment, was not returned for evaluation and the reported event could not be confirmed.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation warranted for this complaint, however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SEMIEXTENDED DRILL GUIDE
Type of Device
ROD, FIXATION, INTRAMEDULLARY AND ACCESSORIES
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key11832854
MDR Text Key250885576
Report Number1020279-2021-04362
Device Sequence Number1
Product Code HSB
UDI-Device Identifier03596010644596
UDI-Public03596010644596
Combination Product (y/n)N
PMA/PMN Number
K122170
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 07/19/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/17/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71654524
Device Catalogue Number71654524
Was Device Available for Evaluation? Yes
Date Manufacturer Received07/16/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-