• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US MBT REV PRIMARY CONICAL REAMER; KNEE INSTRUMENT : REAMERS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US MBT REV PRIMARY CONICAL REAMER; KNEE INSTRUMENT : REAMERS Back to Search Results
Model Number 2178-63-199
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Insufficient Information (4580)
Date of Event 05/04/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint #:(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
It was reported that the two reamer chucks stripped our and the cone reamer is stripped.Needs replaced.Delayed progress of case because it kept stripping.A surgical delay of 10 minutes was noted.
 
Additional Manufacturer Narrative
Product complaint #
=
> (b)(4) this report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d4 (lot number) and d9 if information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected h3.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary =examination of the returned device confirmed the reamer was stripped.The investigation did not indicate that a broader investigation or corrective action was necessary.Depuy considers the investigation closed at this time.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.Device history lot = the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MBT REV PRIMARY CONICAL REAMER
Common Device Name
KNEE INSTRUMENT : REAMERS
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11833206
Report Number1818910-2021-10476
Device Sequence Number17074880
Product Code HTO
UDI-Device Identifier10603295096535
UDI-Public10603295096535
Combination Product (Y/N)N
Initial Reporter StateMO
Initial Reporter CountryUS
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/04/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number2178-63-199
Device Catalogue Number217863199
Device Lot NumberPG215182
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/14/2021
Initial Date Received by Manufacturer 05/04/2021
Supplement Date Received by Manufacturer06/14/2021
07/01/2021
Initial Report FDA Received Date05/17/2021
Supplement Report FDA Received Date07/01/2021
07/02/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-