| Model Number |
400SMTSFT0715 |
| Medical Device Problem Codes |
Failure to Advance (2524); Physical Resistance/Sticking (4012)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/20/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
|
| |
|
Event or Problem Description
|
|
The patient was undergoing a coil embolization procedure in the middle cerebral artery (mca) using penumbra smart coils (smart coil) and a non-penumbra microcatheter.During the procedure, a smart coil was stuck in the starting position and would not advance out of its introducer sheath.Therefore, the smart coil was removed.The procedure was completed using additional smart coils and the same microcatheter.There was no report of an adverse effect to the patient.
|
| |
|
Additional Manufacturer Narrative
|
|
Evaluation of the returned smart coil revealed that the pusher assembly mid-joint was retracted into the introducer sheath friction lock.If this occurs, resistance may be experienced during advancement.During functional testing, the pusher assembly mid-joint was advanced through the introducer sheath friction lock with resistance, then, the smart coil was able to advance through the introducer sheath without an issue.Penumbra coils are inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.
|
| |
|