Model Number 97712 |
Device Problems
Migration or Expulsion of Device (1395); Energy Output Problem (1431); Patient Device Interaction Problem (4001)
|
Patient Problem
Device Overstimulation of Tissue (1991)
|
Event Date 03/14/2017 |
Event Type
Injury
|
Manufacturer Narrative
|
Other relevant device(s) are: product id: neu_unknown_lead, serial/lot #: unknown.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Information was received from multiple sources (healthcare provider, foreign, clinical study) regarding a patient who was implanted with an implantable neurostimulator (ins) for unknown indications for use.It was reported that the patient felt overstimulated when she was sitting or lying down.Lead migration was discovered upon examination on (b)(6) 2017.Fixation of the lead to the osseous plane of the sacrum was done on (b)(6) 2017 and the issue resolved without sequelae.The event resulted in an unscheduled clinic or office visit.The etiology was unlikely related to the device or therapy, and not related to the implant procedure.The device diagnosis was lead migration/dislodgement.It was noted that the patient was (b)(6) years old at consent and weighed (b)(6) kilograms.
|
|
Event Description
|
Additional information received from the manufacturing representative of a clinical study reported that the clinical diagnosis was that the patient felt overstimulation in the lower part of the body.
|
|
Manufacturer Narrative
|
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Additional information was received from the health care provider (hcp) of a clinical study reporting the patient had two leads that migrated.
|
|
Manufacturer Narrative
|
Continuation of d10: product id: 388933; lot#: 0211629779; implanted: on (b)(6) 2016; product type: lead; product id: 388933; lot#: 0211629778; implanted: on (b)(6) 2016; product type: lead.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
|
|
Search Alerts/Recalls
|