| Model Number |
EM100-A |
| Medical Device Problem Codes |
Overheating of Device (1437); No Apparent Adverse Event (3189)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Report confirmed.The device was tested and the max temperature was measured to be 320°f.The likely cause was identified as loose closure collet and shaft assembly.The device user manual warnings section includes instructions that heavy side loads and/or long operating times may cause the device to overheat.If the device overheats, discontinue use and rest the device by using it intermittently, or wrap the device in a moist sterile towel.Use of an overheated device may cause injury to the patient or operator.We will continue to track and trend this complaint type.If information is provided in the future, a supplemental report will be issued.
|
| |
|
Event or Problem Description
|
|
Repair request initiated for device with no reported malfunction.No patient impact reported.Repair escalated to product event on evaluation due to overheating.On follow-up, there was no further information regarding the event and patient or staff impact.
|
| |
|