The certas valve was returned for evaluation: device history record (dhr) - lot number 5101887 with no anomalies.Failure analysis - the position of the cam when valve was received was at setting 3.The valve was visually inspected; the needle guard was raised.A needle hole in the needle chamber was also noted.The valve was hydrated.The valve was leak tested and only leaked from the needle hole in the needle chamber.The valve passed the test for programming, occlusion, reflux and pressure.No root cause could be determined as the technician could not confirm any problem with the valve at the time of investigation.The possible root cause for the issue reported by the customer "once the ventricular catheters was attached they noticed that there was not distal flow" is probably due to wrong handling.As noted in the ifu: "do not fold or bend the valve, folding or bending might cause rupture of the silicone housing, needle guard disc dislodgement or occlusion of the fluid pathway", but at the time of the investigation no occlusion was noted.
|