It was reported that after procedure, was noticed that the black bumper tip that connects to the polarstem ball head impactor on was found too stiff to remove on two devices after use for cleaning (b)(4).There was no patient involved since the event was noticed after procedure.Two polarstem modular head for 21000662 (b)(4) were received, it is unknown what type of failure do these devices have.
|
|
H3, h6: it was reported that after procedure the tip that connects to the polarstem head impactor was found too stiff to remove after use for cleaning.The reported impactor handle (75023710) and the modular head (75023711) , used in treatment, were returned for investigation.It can be confirmed that the parts cannot be disassembled using the usual amount of force.However, the parts could be separated by hand.The impactor handle was heavily deformed and showed scratches which might originate from the use of pliers.The modular head does not show significant signs of use.Therefore, the relationship between devices and the reported event can be confirmed.The production documentation was reviewed.There were no indications found, that the part failed to match specification at the time of manufacturing.For the batches in scope, no further complaints with the same failure mode were reported.The parts were stuck together, most likely because the impactor handle was deformed at the connection site.It must be noted that pliers should never be used to work on the instrument.Therefore, the root cause is attributed to an off-label use.Regarding the modular head, no problems could be found.This complaint was reopened as the parts were returned for investigation.Smith and nephew will monitor this complaint for further similar issues.The returned devices will be discarded.
|