Model Number KD-VC611Q-07301S
Medical Device Problem Code Break (1069)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 01/07/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
The subject device was returned to olympus¿s local distributor, but not returned to omsc.Therefore the exact cause of the reported event could not be conclusively determined at this time.A supplemental report will be submitted, if additional or significant information becomes available at a later time.
 
Event or Problem Description
Olympus medical systems corp.(omsc) was informed by the health professional that during an endoscopic sphincterotomy (est) using the subject device, the cutting wire was broken.There were no missing parts.The intended procedure was completed with another device.There was no patient injury reported.
 
Additional Manufacturer Narrative
This is a supplemental report to provide additional information.The subject device was not returned to olympus medical systems corp.(omsc) for evaluation, because the subject device was discarded by the user.The manufacturing record was reviewed and found no irregularities.It is possible to infer the cause from the results of past similar investigations, therefore it was determined that an investigation using equipment with similar structure or similar device is unnecessary.Based on the past similar cases, omsc presumes that the event occurred due to the following occurrence mechanism.High frequency current was conducted between the cutting wire and the tissue at the point of contact, or high frequency current was conducted when they were being close to each other.Hight frequency current was conducted between the cutting wire and the distal end of the endoscope at the point of contact, or high frequency current was conducted when they are being close to each other.Due to raising the forceps elevator, the cutting wire at the torn area of the coated portion of the wire and the distal end of the endoscope came into contact.Under these circumstances, high frequency current was conducted.If the reported event had occurred in the case of description ¿c¿, a likely mechanism causing the tear of the coated portion of the wire might be the following: the slider was pushed more than needed causing the cutting wire to deflect.The deflected coated portion of the cutting wire and the metal part of the distal end of the endoscope came into contact when the forceps elevator was raised.The tube was moved back and forth in the situation of description ¿2¿, causing the metal part of the distal end of the endoscope to scratch the coated portion of the wire.As a result, the coated portion of the wire was ruptured.The above device handling has warned in the instruction manual.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SINGLE USE PRELOADED SPHINCTEROTOME V (DISTAL WIREGUIDED)
Common Device Name
SINGLE USE PRELOADED SPHINCTEROTOME
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
MDR Report Key11838944
Report Number8010047-2021-06432
Device Sequence Number14397249
Product Code KNS
UDI-Device Identifier04953170399404
UDI-Public04953170399404
Combination Product (Y/N)N
Initial Reporter CountryFR
PMA/510(K) Number
K141991
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional,u
Initial Reporter Occupation Pharmacist
Type of Report Initial,Followup
Report Date (Section B) 06/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberKD-VC611Q-07301S
Device Lot Number05V
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 04/27/2021
Supplement Date Received by Manufacturer06/14/2021
Initial Report FDA Received Date05/17/2021
Supplement Report FDA Received Date06/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-