MAUDE Adverse Event Report: CONCORD MANUFACTURING 2008T HD SYS. CDX BLUE STAR; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

CONCORD MANUFACTURING 2008T HD SYS. CDX BLUE STAR; DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM Back to Search Results
Model Number 191124
Medical Device Problem Code Inadequate Ultra Filtration (1656)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/01/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Plant investigation: no parts were returned to the manufacturer for physical evaluation and an on-site evaluation was not performed by a fresenius field service technician (fst).Additionally, a serial number for the machine could not be provided which prevented the manufacturer from performing a product history review.The investigation into the cause of the reported problem was not able to be confirmed.A definitive conclusion regarding the complaint incident cannot be reached without a physical examination of the complaint device.
 
Event or Problem Description
A user facility manager reported that the ultrafiltration (uf) goal and treatment time changed on a fresenius 2008t hemodialysis (hd) machine after the initiation of the patient's treatment.The machine was programed to a uf goal of 800 ml and treatment time of 3 hours, however, after initiation it was noted that the machine defaulted to a uf rate of 70 ml/hour.The clinical manager stated that the machine could not manually be adjusted to what was programmed by the staff.The uf was not turned off at all during treatment.At that point, treatment was terminated.There was no patient injury, adverse events, or medical intervention required as a result of the reported event.The patient did not have any complaints or symptoms related to the reported event and no additional treatments were required.Per the clinical manager, the reported event could not be duplicated during testing and the machine was returned to service and no parts were replaced or machine repairs made.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
2008T HD SYS. CDX BLUE STAR
Common Device Name
DIALYZER, HIGH PERMEABILITY WITH OR WITHOUT SEALED DIALYSATE SYSTEM
Manufacturer (Section D)
CONCORD MANUFACTURING
director, quality systems
4040 nelson avenue
concord CA 94520
MDR Report Key11839001
Report Number2937457-2021-01062
Device Sequence Number14588907
Product Code KDI
UDI-Device Identifier00840861102082
UDI-Public00840861102082
Combination Product (Y/N)N
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K173972
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source user facility
Initial Reporter Occupation Administrator/Supervisor
Type of Report Initial
Report Date (Section B) 05/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number191124
Device Catalogue Number191124
Was Device Available for Evaluation? No
Device AgeMO
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2021
Initial Report FDA Received Date05/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
FRESENIUS DIALYZER
Patient Age11 MO
Patient Weight9
-
-