|
H10: manufacturing review: a manufacturing review was not required as this is the only complaint reported to date for this product and lot.Investigation summary: one 6s crosser cto recanalization catheter was returned for evaluation.It was noted the distal tip was detached from the catheter and was not returned with sample.The distal marker band was not attached, outer catheter is also noted to be stretched.Distal tip detachment measures 105.5 cm from break to strain relief.Based on the findings, the investigation is confirmed for the reported detachment issue as the it was noted hat the distal tip was completely detached form the catheter.A definitive root cause could not be determined based upon the provided information.Labeling review: the review of the instructions for use, indications, warnings, precautions, cautions, possible complications, and contraindications showed that the product labeling is adequate.H10: d4 (expiration date: 10/2022),g3, h6(method).H11: b5, h1, h6(patient, result and conclusion).H11: section a through f: the information provide by bd represents all the known information at this time.Despite good faith efforts to obtain additional information, the complainant/reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : see h10.
|