• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC PUERTO RICO OPERATIONS CO. PUMP MMT-1715KM 630G 3ML BLACK MEDI  ; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC PUERTO RICO OPERATIONS CO. PUMP MMT-1715KM 630G 3ML BLACK MEDI  ; ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND Back to Search Results
Model Number MMT-1715KM
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Hyperglycemia (1905); Hypoglycemia (1912); Loss of consciousness (2418)
Event Date 05/10/2021
Event Type  Injury  
Manufacturer Narrative
(b)(4).Currently it is unknown whether or not the device may have caused or contributed to the event as no product has been returned.The device will be returned for analysis and further information will follow once the analysis has been completed.No conclusion can be drawn at this time.
 
Event Description
The customer reported via phone call that they were hospitalized due to hypoglycemia on (b)(6) 2021.The customer's blood glucose level at the time of incident was 44 mg/dl.The customer was treated with juice.The customer was passed out.The customer experienced unexpected high and low reading of blood glucose.Customer had been using insulin pump system within 48 hours of reported low blood glucose event.The insulin pump will be returned for analysis.
 
Manufacturer Narrative
Retainer ring = clear on (b)(6) 2021 the customer alleged was hospitalized for high and low blood glucose.The test p-cap locks properly in place in the reservoir compartment noted.The insulin pump received with pillowing keypad overlay and serial number label sticker during the visual inspection.Thus and carelink software was utilized and downloaded trace/history files properly.The insulin pump passed the self test, sleep current measurement, active current measurement, rewind test, prime/seating test, basic occlusion test, occlusion test, force sensor test, displacement test and delivery accuracy test at 0.08710 inches.The insulin pump was cut open to perform visual inspection and found no moisture or component damage on the electronics, force sensor and motor assembly noted.The force sensor offset measured (22.1 milli volts).The motor was tested outside of the device on the ngp stb3 and passed.In summary, insulin pump passed all required testing.Unable to verify customer alleged for high and low blood glucose.The force sensor is within specification and the motor functioning properly.This mdr related to the puerto rico manufacturing site has been assigned a medwatch number from the medtronic minimed northridge site, per variance 5.Medtronic, inc.(medtronic) is submitting this report to comply with 21 c.F.R.Part 803, the medical device reporting regulation.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information in the time allotted and has provided as much information as is available to the company as of the submission date this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employees caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any "defects" or has "malfunctioned".These words are included in the fda 3500a form and are fixed items for selection created by the fda, to categorize the type of event solely for the purpose of reporting pursuant to part 803.Medtronic objects to the use of these words and others like it because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PUMP MMT-1715KM 630G 3ML BLACK MEDI  
Type of Device
ARTIFICIAL PANCREAS DEVICE SYSTEM, THRESHOLD SUSPEND
Manufacturer (Section D)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer (Section G)
MEDTRONIC PUERTO RICO OPERATIONS CO.
ceiba norte ind. park #50 road
juncos 00777 -386
Manufacturer Contact
tricha miles
ceiba norte ind. park #50 road
juncos 00777--386
7635140379
MDR Report Key11840060
MDR Text Key251098593
Report Number2032227-2021-146381
Device Sequence Number1
Product Code OZO
UDI-Device Identifier000000643169873834
UDI-Public(01)000000643169873834
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Consumer
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 02/16/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2021
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Lay User/Patient
Device Model NumberMMT-1715KM
Device Catalogue NumberMMT-1715KM
Device Lot NumberHG206J7
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/09/2021
Is the Reporter a Health Professional? No
Date Manufacturer Received02/10/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured08/14/2017
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
Treatment
FRN-UNK-RSVR, UNOMED SET; FRN-UNK-RSVR, UNOMED SET
Patient Outcome(s) Hospitalization;
Patient Age73 YR
Patient SexFemale
-
-