Model Number 363430 |
Device Problem
Fluid/Blood Leak (1250)
|
Patient Problem
Hemorrhage/Bleeding (1888)
|
Event Date 04/22/2021 |
Event Type
malfunction
|
Manufacturer Narrative
|
This report has been identified as b.Braun medical internal report number (b)(4).The investigation is ongoing at this time.A follow up will be submitted when the investigation results become available.
|
|
Event Description
|
As reported by the user facility: the patient did not receive the proper treatment which is the propofol sedation required.Luckily it did not cause any complication in the patient condition.On the other hand the patient lost blood from the leaking port.This was an unnecessary blood lost.Since this is an opened medium, this puts the patient at risk for infection.According to the customer, the product leaked from the part of insertion of the needle causing injuries to the patient.The central line blood backing up and leaking through a primary line port.The primary line had ns running and the port at which blood leaked had propofol line attached to it.The port had a hole.
|
|
Manufacturer Narrative
|
This report has been identified as b.Braun medical internal report number (b)(4).No sample was provided for evaluation.Further investigation of the complaint is not possible.Review of the discrepancy management system (dsms) database was performed for the reported lot number and no abnormalities or non-conformances were noted during the in process or final product inspection.We will maintain this report for further references and continue to monitor other reports for similar occurrences.If any additional pertinent information becomes available, a follow up will be submitted.
|
|