• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ACCU-PASS STR SHUTTLE 45 DEG LFT CURVE; TROUSERS, ANTI-SHOCK

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ACCU-PASS STR SHUTTLE 45 DEG LFT CURVE; TROUSERS, ANTI-SHOCK Back to Search Results
Model Number 7210423
Medical Device Problem Code Mechanical Problem (1384)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Internal complaint reference (b)(4).
 
Event or Problem Description
It was reported that during a shoulder arthroscopy case, the accu-pass monofilament suture was not passing smoothly and the wheels in the device are difficult to use.The procedure was successfully completed without a significant delay using a back-up device.No patient injury or other complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: (b)(4).The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was a repeat issue.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCU-PASS STR SHUTTLE 45 DEG LFT CURVE
Common Device Name
TROUSERS, ANTI-SHOCK
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key11841054
Report Number1219602-2021-01155
Device Sequence Number14367657
Product Code LHX
UDI-Device Identifier03596010534446
UDI-Public03596010534446
Combination Product (Y/N)N
Initial Reporter CountryHK
PMA/510(K) Number
EXEMPT
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source foreign,health professional,o
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date09/19/2023
Device Model Number7210423
Device Catalogue Number7210423
Device Lot Number2061337
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/27/2021
Supplement Date Received by Manufacturer06/18/2021
Initial Report FDA Received Date05/18/2021
Supplement Report FDA Received Date06/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-