| Model Number |
98814-65 |
| Medical Device Problem Code |
Incorrect, Inadequate or Imprecise Result or Readings (1535)
|
| Health Effect - Clinical Code |
Insufficient Information (4580)
|
| Date of Event |
04/27/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
No product has been returned, extended investigation is pending at this time.A follow- up report will be submitted once additional information is obtained.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter.Customer reported receiving readings of 69 mg/dl, 485 mg/dl and 125 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|
Additional Manufacturer Narrative
|
|
No product has been returned.Extended investigation has been performed for the reported complaint and there was no indication that the product did not meet specification.Pqe reviewed the dhr for the xceed meter indicated by the serial number provided by the customer.The dhr showed the xceed meter passed all tests prior to release.Pqe reviewed the dhr for the precision strips indicated by the serial number provided by the customer.The dhr showed the precision strips passed all tests prior to release.Retain testing was performed for precision strips and all units performed in specification and passed.If the product is returned, a physical investigation will be performed, and a follow-up report submitted.All pertinent information available to abbott diabetes care has been submitted.
|
| |
|
Event or Problem Description
|
|
Customer reported receiving erratic readings on their adc blood glucose meter.Customer reported receiving readings of 69 mg/dl, 485 mg/dl and 125 mg/dl within 10 minutes.The results when plotted on a parkes error grid fell into the "c" zone showing the difference in values to be clinically significant.There was no report of death, serious injury, or mistreatment associated with this event.
|
| |
|