• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDOS INTERNATIONAL SARL MICRUSFRAME18 11MM X 37CM; NEUROVASCULAR EMBOLIZATION DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDOS INTERNATIONAL SARL MICRUSFRAME18 11MM X 37CM; NEUROVASCULAR EMBOLIZATION DEVICE Back to Search Results
Model Number MFR181137
Medical Device Problem Codes Stretched (1601); Failure to Advance (2524)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/27/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Product complaint # (b)(4).Information regarding patient weight, height, medical history, race, and ethnicity was not reported.Procode: krd/hcg.The device is available to be returned for evaluation and testing.However, it has not been received to date as indicated as ¿other¿ in this section as the reason for non-evaluation.If the device returns, a device investigation will be performed.A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Event or Problem Description
As reported by the field, during a coil embolization, a 11mm x 37cm micrusframe18 coil (mfr181137, 30415698) was inserted correctly, but already with difficulty.The coil reached the aneurysm, then could not be pushed further.When pulled back, the coil stretched just before the end.There was no patient injury reported.
 
Additional Manufacturer Narrative
Product complaint (b)(4).A supplemental report will be submitted if new facts arise which materially alter information submitted in a previous mdr report.Complaint conclusion: as reported by the field, during a coil embolization, a 11mm x 37cm micrusframe18 coil (mfr181137, 30415698) was inserted correctly, but already with difficulty.The coil reached the aneurysm, then could not be pushed further.When pulled back, the coil stretched just before the end.There was no patient injury reported.No additional information is available.The device was discarded; therefore, no further investigation can be performed.A manufacturing record evaluation (mre) was performed for the finished device 30415698 number, and no non-conformances related to the reported complaint condition were identified.With the information available and without the product available for analysis, the reported customer complaints of ¿detachable coil delivery system (dcs) - impeded in microcatheter with no loss of cerebral target position¿ and ¿coil - unraveled/stretched-in microcatheter¿ could not be confirmed.Based on the manufacturing record evaluation, there is no indication that the events are related to the device manufacturing process.Impeded in microcatheter is a known potential product failure associated with the use of the device.The ifu includes warnings and instructions for use to trouble shoot this type of situation.The instructions for use (ifu) warns that if the microcoil system becomes immobile in the infusion microcatheter, apply a gentle push-pull motion to free it.If unsuccessful, remove both microcatheter and microcoil system together as a unit and replace with new devices.Assignment of root cause for the events remains speculative and inconclusive, based on the limited information provided and without the return of the complaint device; however, it is possible that clinical and procedural factors, including device manipulation / interaction, aneurysm size and vessel characteristics, device selection, and the concomitant microcatheter, may have contributed to the reported issues.An embolic coil becomes stretched when an excessive pull force is applied to the device.As part of the post market surveillance program, information from this complaint is trended to identify statistical signals for consideration of further correction action.Since there was no evidence to suggest the events were related to a manufacturing or design issue, no corrective actions will be taken at this time.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
MICRUSFRAME18 11MM X 37CM
Common Device Name
NEUROVASCULAR EMBOLIZATION DEVICE
Manufacturer (Section D)
MEDOS INTERNATIONAL SARL
chemin-blanc 38
le locle neuchatel CH-24 00
SZ  CH-2400
MDR Report Key11841423
Report Number3008114965-2021-00199
Device Sequence Number19748000
Product Code KRD
UDI-Device Identifier10886704078074
UDI-Public10886704078074
Combination Product (Y/N)N
PMA/510(K) Number
K150319
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source foreign,health professional
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 04/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date06/30/2023
Device Model NumberMFR181137
Device Catalogue NumberMFR181137
Device Lot Number30415698
Was Device Available for Evaluation? No
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/28/2021
Supplement Date Received by Manufacturer06/08/2021
Initial Report FDA Received Date05/18/2021
Supplement Report FDA Received Date06/17/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-