SMITH & NEPHEW, INC. LGN POROUS CR FEM SZ 7L; PROSTHESIS, KNEE, PATELLO/FEMOROTIBIAL, SEMI-CONSTRAINED, UNCEMENTED, POROUS, CO
|
Back to Search Results |
|
| Model Number |
71423247 |
| Medical Device Problem Code |
Expiration Date Error (2528)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that after tka surgery, it was found out that the patient could have possibly received a lgn porous cr fem sz 7l expired.However, the labeling of the implant contains the wrong expiration date.It was use the same s+n backup device.No patient harm or other complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The associated device, used in treatment, was not returned for evaluation but the reported event could be confirmed through a review of the manufacturing records.A review of the manufacturing records did reveal that the expiration date on the router is 08/2021 but the expiration date on the labels is 12/2020.A review of complaint history did not reveal additional complaints for the listed batch.At this time, we have reason to suspect that the product failed to meet product specifications at the time of manufacture.The potential probable cause for this event is a label system error.This issue was evaluated through our internal quality hold process which found that previous actions have already been taken to address this failure.A new labeling system has been implemented since this event and will eliminate the reoccurrence of this failure.Based on this investigation, the need for corrective action is not indicated.Without the return of the actual product involved, our investigation could not proceed.Should the device or additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.
|
| |
|
Search Alerts/Recalls
|
|
|