• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: PENUMBRA, INC. LANTERN DELIVERY MICROCATHETER; DQY

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

PENUMBRA, INC. LANTERN DELIVERY MICROCATHETER; DQY Back to Search Results
Model Number PXSLIMLAN150STR
Medical Device Problem Codes Break (1069); Physical Resistance/Sticking (4012)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
This device is available for return.A follow up mdr will be submitted upon completion of the device investigation.
 
Event or Problem Description
The patient was undergoing a medical procedure in the internal iliac/hypogastric artery using a lantern delivery microcatheter (lantern), a base catheter, a 6f sheath, and a guidewire.During the procedure, the physician advanced the lantern over the guidewire into the target vessel.The physician then performed a few power injections through the lantern to visualize the vessel.Next, the physician attempted to advance the lantern further.While re-inserting the guidewire into the lantern, the physician experienced resistance, and subsequently, the lantern would not advance further.Therefore, the physician decided to remove the lantern and the guidewire.However, while removing the lantern, the physician experienced resistance and pulled the lantern with force.Upon removal of the entire lantern, the physician noticed that the lantern was broken at the distal end.The procedure was completed using another catheter.There was no report of an adverse effect to the patient.
 
Additional Manufacturer Narrative
Evaluation of the returned lantern confirmed that the catheter was fractured.Fracture typically occurs due to manipulation against resistance.Based on the reported event, the catheter was being manipulated against resistance within the patient anatomy while re-inserting a guidewire and upon removal of the lantern.This likely contributed to the fracture during the procedure.The root cause of resistance could not be determined.Penumbra devices are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
LANTERN DELIVERY MICROCATHETER
Common Device Name
DQY
Manufacturer (Section D)
PENUMBRA, INC.
one penumbra place
alameda CA 94502
MDR Report Key11842395
Report Number3005168196-2021-01076
Device Sequence Number3660917
Product Code DQY
UDI-Device Identifier00814548016672
UDI-Public00814548016672
Combination Product (Y/N)Y
Initial Reporter StateNV
Initial Reporter CountryUS
PMA/510(K) Number
K152840
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 04/19/2021
Report Date (Section F) 01/01/2005
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date11/05/2022
Device Model NumberPXSLIMLAN150STR
Device Catalogue NumberPXSLIMLAN150STR
Device Lot NumberF94074
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/15/2021
Was the Report Sent to FDA? Yes
Date Report to Manufacturer01/10/2005
Initial Date Received by Manufacturer 04/19/2021
Supplement Date Received by Manufacturer06/30/2021
Initial Report FDA Received Date05/18/2021
Supplement Report FDA Received Date07/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age85 YR
Date Report Sent to FDA01/01/2005
-
-