The patient was undergoing a medical procedure in the internal iliac/hypogastric artery using a lantern delivery microcatheter (lantern), a base catheter, a 6f sheath, and a guidewire.During the procedure, the physician advanced the lantern over the guidewire into the target vessel.The physician then performed a few power injections through the lantern to visualize the vessel.Next, the physician attempted to advance the lantern further.While re-inserting the guidewire into the lantern, the physician experienced resistance, and subsequently, the lantern would not advance further.Therefore, the physician decided to remove the lantern and the guidewire.However, while removing the lantern, the physician experienced resistance and pulled the lantern with force.Upon removal of the entire lantern, the physician noticed that the lantern was broken at the distal end.The procedure was completed using another catheter.There was no report of an adverse effect to the patient.
|
Evaluation of the returned lantern confirmed that the catheter was fractured.Fracture typically occurs due to manipulation against resistance.Based on the reported event, the catheter was being manipulated against resistance within the patient anatomy while re-inserting a guidewire and upon removal of the lantern.This likely contributed to the fracture during the procedure.The root cause of resistance could not be determined.Penumbra devices are visually inspected during in-process inspection and during quality inspection after manufacturing.The manufacturing records for this lot were reviewed and did not reveal any outstanding discrepancies, design, or quality concerns.H3 other text : placeholder.
|