The reported events of lead pin not fully inserting into the connector adaptor was confirmed.As received, a complete lead was returned for evaluation, measuring 75.3 cm.Electrical tests did not find any anomalies.The lead connector passed insertion testing into the test implantable cardioverter defibrillator device, but failed insertion testing into the test is-4 connector sleeve.Dimensional analysis of the connector identified an over-sized diameter in a section of the connector boot.This may have contributed to the reported field event.
|