| Model Number |
QD8-S |
| Medical Device Problem Code |
Vibration (1674)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The device was returned for evaluation.During repair, an evaluation was performed and it was determined that the reported condition was confirmed.The assignable root cause was traced to component failure due to normal wear.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported from (b)(6) that during service and evaluation, it was determined that the attachment device was vibrating and the cutter device could not be inserted.It was further determined that the device failed pretest for cutter insertion and vibration.It was noted in the service order that the device was out of order.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|