Model Number PM438R
Device Problem Material Fragmentation (1261)
Patient Problem Burn(s) (1757)
Event Date 03/27/2021
Event Type  malfunction  
Manufacturer Narrative
Manufacturing site evaluation: investigation on-going.Additional information / investigation results will be provided in a supplemental report.
 
Event Description
It was reported that there was an issue with pm438r - jaw ins.Bip.Maryland diss.Fen.5/310mm.According to the complaint description, the a risk of untimely coagualtion occured during surgery.Involuntary coagulation of the small intestine by the metal part upstream of the joint of the bipolar forceps during a laparoscopic appendectomy.A burn occured.Additional information was not provided.Additional patient information is not available.The malfunction is filed under aag reference (b)(4).Associated medwatch-reports: 9610612-2021-00380 ((b)(4)+ pm438r) 9610612-2021-00381 ((b)(4)+ pm438r) 9610612-2021-00382 ((b)(4)+ pm438r) 9610612-2021-00383 ((b)(4)+ pm438r).
 
Manufacturer Narrative
Investigation results: visual investigation: the investigation was carried out visually and microscopically.Visual inspection was carried out for all of the products.The products shows these or similar deviations and / or damages.Here we found a visible damaged and broken blue ceramic.We also detected a visible damaged and broken grey ceramic.Additionally we discovered misaligned jaws.Investigations lead to the assumption that the ceramic breakages and the misaligned jaw parts caused by an improper handling due to a mechanical overload situation.Possibly an excessive force has been applied on the instrument or the possibility of torsion or high leverage with the instrument.A major disadvantage of ceramics is their brittle fracture behavior (limited elongation and low fracture toughness).Metallic materials, on the other hand, are ductile and therefore break less frequently.They forgive easier constructive tolerances by relieving local stress peaks through elastic and plastic deformation.A damage isolation or ceramic could lead to an electrical malfunction.Batch history review: the device quality and manufacturing history records have been checked for the available lot number and were found to be according to our specifications valid at the time of production.Review of the complaint history revealed that no similar complaints have been filed against products from this batch number.The review of risk assessment revealed that the overall risk level (severity 4(5) x probability of occurrence 1(5)) according to din en iso 14971 is still acceptable.Conclusion and measures / preventive measures: based upon the investigation results a clear root cause conclusion cannot be drawn.There is no indication for a material-, manufacturing- or design-related failure.Based upon the investigations results a product safety case (psc 2021-039) was internally initiated for further investigations.
 
Event Description
Associated medwatch-reports: 9610612-2021-00381 (400511104 + pm438r), 9610612-2021-00382 (400511105 + pm438r), 9610612-2021-00383 (400511106 + pm438r).
 
Manufacturer Narrative
Associated medwatch-reports: 9610612-2021-00380 (400511103 + pm438r) 9610612-2021-00381 (400511104 + pm438r) 9610612-2021-00382 (400511105 + pm438r) 9610612-2021-00383 (400511106 + pm438r) after reassessment this case was found to be non-reportable.The failure mode has changed, and the failure risk analysis adjusted from 4(5) severity to 2(5) severity.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
JAW INS.BIP.MARYLAND DISS.FEN.5/310MM
Type of Device
BIPOLAR REUSABLE INSTRUMENTS
Manufacturer (Section D)
AESCULAP AG
po box 40
tuttlingen, 78501
GM  78501
Manufacturer (Section G)
AESCULAP AG
po box 40
tuttlingen, 78501
GM   78501
Manufacturer Contact
christian von der grün
po box 40
tuttlingen, 78501
GM   78501
MDR Report Key11842953
MDR Text Key251412543
Report Number9610612-2021-00380
Device Sequence Number1
Product Code HGI
Combination Product (y/n)N
Reporter Country CodeFR
PMA/PMN Number
K003608
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Foreign,Company Representative
Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup,Followup
Report Date 03/03/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberPM438R
Device Catalogue NumberPM438R
Device Lot Number62429470
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2021
Is the Reporter a Health Professional? Yes
Date Manufacturer Received02/14/2022
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured12/21/2020
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-