Model Number 0502539010
Medical Device Problem Code Optical Problem (3001)
Health Effect - Clinical Codes Insufficient Information (4580); No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/19/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
Additional information will be provided once the investigation has been completed.The device manufacturer date is not known at this time.However, should it become available it will be provided in future reports.
 
Event or Problem Description
It was reported that there was foggy image.
 
Additional Manufacturer Narrative
Alleged failure: foggy confirmed failure: outer tube damaged (bent, dented),moisture ingression at distal end,moisture ingression at proximal end,broken rod lens.Probable root cause: laser welding seal failure.Distal/proximal window solder failure.Damage to optical train.Damage to needle.Damage to distal or proximal windows.Moisture intrusion.End of life wear-out.Environmental disturbance: endoscope colder than dew point.Environmental disturbance: fluid entrapment between the coupler and eyepiece junction (eyepiece only).Use error.The product was returned for investigation and the reported failure mode was confirmed.The failure mode will be monitored for future reoccurrence.The manufacture date is not known.
 
Event or Problem Description
It was reported that there was foggy image.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HD 5MM 0A LAPAROSCOPE, AC
Common Device Name
LAPAROSCOPE, GENERAL & PLASTIC SURGERY
Manufacturer (Section D)
STRYKER ENDOSCOPY-SAN JOSE
5900 optical court
san jose CA 95138
MDR Report Key11843382
Report Number0002936485-2021-00262
Device Sequence Number4395721
Product Code GCJ
UDI-Device Identifier07613327059403
UDI-Public07613327059403
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
K910132
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source consumer
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 07/07/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model Number0502539010
Device Catalogue Number0502539010
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer04/27/2021
Initial Date Received by Manufacturer 04/19/2021
Supplement Date Received by Manufacturer04/19/2021
Initial Report FDA Received Date05/18/2021
Supplement Report FDA Received Date07/07/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-