Model Number 75100444
Medical Device Problem Code Mechanical Jam (2983)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that during a thr, the plug cover of a polarcup shell ti-plasma/ha 59 non-cem couldn't be removed and was worn out while being unplugged.There was a delay of 30 minutes or less and there was a change of surgical procedure due to the failure.Patient was not harmed.
 
Additional Manufacturer Narrative
H3, h6: it was reported that during a thr, the plug cover of a polarcup shell ti-plasma/ha 59 non-cem (75100444) couldn't be removed and was worn out while being unplugged.The device intended for use in treatment was not returned for investigation.An evaluation of the complained device could therefore not be conducted and the reported failure mode could not independently be confirmed.A review of the complaint history revealed no other complaint being reported for the batch in question.A review of the batch record revealed no deviations from the standard manufacturing process that could have contributed to the reported failure mode.The surgical technique (lit no.01620-en (1582) v7 10/20) describes the correct removal of the polarcup plugs using the unidirectional t handle (75023347): "release the plugs by turning the t wrench in the direction indicated on the plug cover." the ifu (lit.No.12.23 ed 05/16) states: "the following might not be implanted under any circumstances: implant components that have been damaged or scratched; implants that have been handled inappropriately or processed in a way that is not part of the surgical technique".The current risk analysis covers the risk of jamming between plug and shell.The associated risk level is considered low.Based on the performed investigations, the reported failure mode could not be confirmed.A relationship between the reported event and the device cannot be confirmed.There is no indication that the reported device failed to meet manufacturing specifications upon release for distribution.A worn t-handle is known to contribute to the reported event.According to document "processing (cleaning, disinfection and sterilization) of instruments from smith & nephew orthopaedics ag" (lit.N°03389-en 1363 v3 11/19), all devices must be inspected and controlled for proper functioning after cleaning/disinfection.Based on available information the root cause for the reported jamming cannot clearly be identified and stays undetermined.The need for corrective action is not indicated.Nevertheless, smith + nephew will continue to monitor this device for similar issues.This complaint will be reopened should additional information or the device be received.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
POLARCUP SHELL TI-PLASMA/HA 59 NON-CEM
Common Device Name
PROSTHESIS, HIP, SEMI-CONSTRAINED, METAL/POLYMER, CEMENTED
Manufacturer (Section D)
SMITH & NEPHEW ORTHOPAEDICS AG
schachenallee 29
aarau CH-50 00
SZ  CH-5000
MDR Report Key11843411
Report Number9613369-2021-00228
Device Sequence Number4545416
Product Code JDI
UDI-Device Identifier07611996118353
UDI-Public07611996118353
Combination Product (Y/N)N
Initial Reporter CountryPK
PMA/510(K) Number
K122244
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source health professional
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 07/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date04/25/2021
Device Model Number75100444
Device Catalogue Number75100444
Device Lot NumberB2006690
Was Device Available for Evaluation? Yes
Initial Date Received by Manufacturer 04/29/2021
Supplement Date Received by Manufacturer07/08/2021
Initial Report FDA Received Date05/18/2021
Supplement Report FDA Received Date07/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Hospitalization; Required Intervention;
-
-