• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION WATCHMAN LAA CLOSURE DEVICE & DELIVERY SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION WATCHMAN LAA CLOSURE DEVICE & DELIVERY SYSTEM; SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL Back to Search Results
Model Number 10371
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Code Pericardial Effusion (3271)
Date of Event 04/29/2021
Type of Reportable Event Serious Injury
Event or Problem Description
It was reported that a pericardial effusion occurred.A left atrial appendage (laa) closure procedure was being performed.A watchman access system (was) was positioned and a 21mm watchman laa closure device & delivery system (wds) were used.The closure device was pushed forward during deployment, which caused a pericardial effusion in the anterior lobe of the laa.The closure device was recaptured, re-deployed and successfully implanted.Pericardiocentesis was performed to treat the pericardial effusion with 650 ml of blood withdrawn from the effusion site.The drainage catheter was left in the patient overnight.The patient was discharged one day post procedure in stable condition.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
WATCHMAN LAA CLOSURE DEVICE & DELIVERY SYSTEM
Common Device Name
SYSTEM, APPENDAGE CLOSURE, LEFT ATRIAL
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
two scimed place
maple grove MN 55311
MDR Report Key11843544
Report Number2134265-2021-06197
Device Sequence Number11619252
Product Code NGV
UDI-Device Identifier08714729838227
UDI-Public08714729838227
Combination Product (Y/N)N
Initial Reporter StateTX
Initial Reporter CountryUS
PMA/510(K) Number
P130013
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial
Report Date (Section B) 05/18/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date12/05/2022
Device Model Number10371
Device Catalogue Number10371
Device Lot Number0024884049
Was Device Available for Evaluation? No
Type of Report(Section G)Initial
Initial Date Received by Manufacturer 04/29/2021
Initial Report FDA Received Date05/18/2021
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-