• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARROW INTERNATIONAL INC. AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARROW INTERNATIONAL INC. AC3 OPTIMUS IABP NA/EMEA; SYSTEM, BALLOON, INTRA-AORTIC Back to Search Results
Model Number IPN001112
Device Problem Gas/Air Leak (2946)
Patient Problem Appropriate Clinical Signs, Symptoms, Conditions Term / Code Not Available (4581)
Event Date 05/03/2021
Event Type  malfunction  
Manufacturer Narrative
(b)(4).
 
Event Description
There was no patient involvement.It was reported that the intra-aortic balloon pump (iabp) have been using "a lot of helium".The user stated that the tanks seem to be empty when the iabp weren't even in use.
 
Manufacturer Narrative
Qn# (b)(4).The product was not returned for investigation.The reported complaint of helium tank running out quickly was confirmed by the field service agent.According to the attached fsr, the pcs assembly was replaced.The root cause of the complaint is undetermined.If the product is returned at a later date, a full investigation of the sample will be completed.A device history record (dhr) review was conducted for the lot number with no relevant findings.The device passed all manufacturing specifications prior to release.Teleflex assessed the risk for the reported complaint.There are no new or revised risk.This will be monitored for any developing trends.
 
Event Description
There was no patient involvement.It was reported that the intra-aortic balloon pump (iabp) have been using "a lot of helium".The user stated that the tanks seem to be empty when the iabp weren't even in use.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
AC3 OPTIMUS IABP NA/EMEA
Type of Device
SYSTEM, BALLOON, INTRA-AORTIC
Manufacturer (Section D)
ARROW INTERNATIONAL INC.
reading PA
MDR Report Key11843873
MDR Text Key252243005
Report Number3010532612-2021-00140
Device Sequence Number1
Product Code DSP
UDI-Device Identifier30801902084966
UDI-Public30801902084966
Combination Product (y/n)N
PMA/PMN Number
K162820
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 05/03/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Date FDA Received05/18/2021
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberIPN001112
Device Catalogue NumberIAP-0700
Device Lot NumberN/A
Was Device Available for Evaluation? Yes
Date Manufacturer Received06/22/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-