| Model Number |
3243 |
| Medical Device Problem Code |
Device Contamination with Chemical or Other Material (2944)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
05/04/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Event or Problem Description
|
|
It was reported that there was foreign material on the burr tip.A 1.75mm rotapro was selected for use in an atherectomy procedure.During loading of the rotapro onto the guidewire, debris was observed on the tip of the burr.The burr never entered the patient's body.The rotapro was removed and exchanged for another of the same.No patient complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Device evaluated by mfr: the returned product consisted of the rotapro atherectomy system.The burr catheter was received detached from the advancer unit.The advancer, handshake connections, sheath, coil, burr and annulus were visually and microscopically examined.Inspection presented no damage or irregularities to the device.The burr was inspected, and no debris was identified on the tip or body of the burr.
|
| |
|
Event or Problem Description
|
|
It was reported that there was foreign material on the burr tip.A 1.75mm rotapro was selected for use in an atherectomy procedure.During loading of the rotapro onto the guidewire, debris was observed on the tip of the burr.The burr never entered the patient's body.The rotapro was removed and exchanged for another of the same.No patient complications were reported.
|
| |
|