| Model Number |
OM-9020 |
| Medical Device Problem Code |
Break (1069)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/29/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: case (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that, during an arthroscopy procedure, the t handle on the magnum punch snapped off whilst pulling out of the patient.The procedure was completed with the same device with no patient injury.A delay equal to or less than 30 minutes was reported.No further complications reported.
|
| |
|
Additional Manufacturer Narrative
|
|
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.Insufficient product identification information was provided and thus a manufacturing record review could not be conducted.A complaint history review concluded this was a repeat issue.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.H6: updated.
|
| |
|