ZIMMER BIOMET, INC. G7 STR MONOBLOCK SHELL INSRTR; INSTRUMENT, HIP Back to Search Results
Model Number N/A
Medical Device Problem Code Fracture (1260)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
(b)(4).Product has been returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.Multiple mdr reports were filed for this event, please see associated reports: 0001825034-2021-01514.
 
Event or Problem Description
It was reported that threaded tip on inserter broke off in the 32mm impactor when surgeon was impacting liner.All parts were retrieved.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Additional Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The following sections were updated/corrected updated: g3; h2; h3; h6.   one g7 str monoblock shell insrtr was returned and evaluated.Upon visual inspection the ball impactor has a square indentation on the od of the device and there are fractured threads inside the threaded portion of the device.The returned inserter has impact marks on the strike plate and some scuffing on the shaft.The threaded tip of the device had fractured from the device.A review of the device history records identified deviations or anomalies during manufacturing, the deviations or anomalies would not have attributed to the event.Medical records were not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event or Problem Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
G7 STR MONOBLOCK SHELL INSRTR
Common Device Name
INSTRUMENT, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key11847716
Report Number0001825034-2021-01513
Device Sequence Number16949777
Product Code PBI
UDI-Device Identifier00880304532694
UDI-Public(01)00880304532694(11)150911(10)252270
Combination Product (Y/N)N
Initial Reporter StateFL
Initial Reporter CountryUS
PMA/510(K) Number
K121874
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source distributor,health profession
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberN/A
Device Catalogue Number110003450
Device Lot Number252270
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/11/2021
Was the Report Sent to FDA? No
Initial Date Received by Manufacturer 05/04/2021
Supplement Date Received by Manufacturer09/07/2021
Initial Report FDA Received Date05/19/2021
Supplement Report FDA Received Date09/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
If action reported to FDA under 21 USC 360i(g), list
FDA-assigned Recall Number or include a statement
N/A
Patient Sequence Number1
-
-