• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY ORTHOPAEDICS INC US S-ROM*SLEEVE PRX ZTT, 16B-LRG; S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY ORTHOPAEDICS INC US S-ROM*SLEEVE PRX ZTT, 16B-LRG; S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE Back to Search Results
Model Number 52-1465
Medical Device Problem Code Adverse Event Without Identified Device or Use Problem (2993)
Health Effect - Clinical Codes Foreign Body Reaction (1868); Pain (1994); Synovitis (2094); Ambulation or Postural Difficulties (2544); Swelling/ Edema (4577); Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 05/21/2019
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Product complaint # (b)(4).Initial reporter occupation: lawyer.(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event or Problem Description
Asr revision litigation records received.Records alleges pain, injury, economic loss, walking difficulty and excessive cobalt and chromium levels.Doi: (b)(6) 2008, dor: (b)(6)2019, right hip.Medical records received.After review of medical records patient was revised to addressed metallosis.Operative notes indicated in mri shows some diffuse synovitis.Upon opening the capsule there was some fluid pressure that appears more milky.There was a lot of staining of the trunnion and a little bit pitting.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.Device history lot: the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Additional Manufacturer Narrative
Product complaint # (b)(4).Investigation summary : no device associated with this report was received for examination.A worldwide lot specific complaint database search, or manufacturing record evaluation, was not possible as the required lot number was not provided.The information received will be retained for trend analysis, post market surveillance, or other events within the quality system.Device history lot : the product investigation found no evidence suspecting an error in the manufacturing or material that would be a contributing factor in the reported allegation(s).A manufacturing records evaluation (mre) was not performed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Common Device Name
S-ROM HIP SYSTEM : HIP FEMORAL SLEEVE
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key11847970
Report Number1818910-2021-10633
Device Sequence Number3224203
Product Code LPH
UDI-Device Identifier10603295170624
UDI-Public10603295170624
Combination Product (Y/N)N
Initial Reporter StateNY
Initial Reporter CountryUS
PMA/510(K) Number
P070026
Number of Events Summarized1
Summary Report (Y/N)N
Device Implanted Year2008
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source other
Initial Reporter Occupation Other
Type of Report Initial,Followup,Followup
Report Date (Section B) 05/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Operator of Device Health Professional
Device Expiration Date07/31/2012
Device Model Number52-1465
Device Catalogue Number521465
Device Lot Number2435382
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/14/2021
Supplement Date Received by Manufacturer05/24/2021
08/13/2021
Initial Report FDA Received Date05/19/2021
Supplement Report FDA Received Date05/25/2021
08/19/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Concomitant Medical Products
and Therapy/Usage Dates
ADAPTER SLEEVE 11/13 +3.; ASR ACETABULAR CUPS 46.; ASR UNI FEMORAL IMPL SIZE 41.; S-ROM*STM STD,36NK,16X11X150.; ADAPTER SLEEVE 11/13 +3; ASR ACETABULAR CUPS 46; ASR UNI FEMORAL IMPL SIZE 41; S-ROM*STM STD,36NK,16X11X150
Outcome Attributed to Adverse Event Required Intervention;
Patient Age68 YR
Patient Weight86
-
-