| Model Number |
CSE-P-125-JA |
| Medical Device Problem Code |
Fluid/Blood Leak (1250)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/11/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation of the returned product and photos confirmed the malfunction of inner crack of the bowl.Although there was no serious injury or harm, past reporting indicates this particular malfunction has been associated with a reported death.The investigation of the past report that had a death indicated the inner core leak malfunctioned but did not cause or contribute to the reported incident according to the surgeon that performed the surgery.Haemonetics decided to conservatively report inner core leak due the confirmed malfunction that occurred.Manufacturing evaluation confirmed inner core crack malfunction.
|
| |
|
Event or Problem Description
|
|
Customer in (b)(6) informed haemonetics the cell saver elite device displayed a long empty error while using a similar device disposable set cell saver elite set - 125ml.There was no report of patient ha.
|
| |
|