| Model Number |
7010 |
| Medical Device Problem Code |
Material Integrity Problem (2978)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/21/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(6).Device evaluated by mfr.: returned product consisted of a quantum maverick balloon catheter.The device was microscopically and visually examined.At 50.2cm from strain relief there was a kink in the hypotube and full separation of device at 53.5cm from strain relief.There was blood in the guidewire lumen.There was contrast and blood in the folds of the balloon and the balloon was tightly folded.The device also had tip damage.Product analysis does not confirm reported allegation of balloon damage to the device as there was no damage found to the balloon.There was unreported damage to the device as there was a separation of the hypotube, a hypotube kink and tip damage.
|
| |
|
Event or Problem Description
|
|
Reportable based on device analysis completed on 07may2021.It was reported that the balloon was damaged.A 4.0mm x 12mm quantum maverick was seleced for use.However, it was noted that the balloon was found damage when unpacked.The procedure was completed with another of the same device.There were no patient complications reported.However, returned device analysis revealed shaft detached/separated.
|
| |
|
Search Alerts/Recalls
|