Model Number FX417T
Medical Device Problem Code Infusion or Flow Problem (2964)
Health Effect - Clinical Codes Head Injury (1879); Hemorrhage/Blood Loss/Bleeding (1888); Hydrocephalus (3272)
Date of Event 04/09/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Additional information/ investigation result will be submitted in a supplemental report.
 
Event or Problem Description
It was reported that there was an issue with fv046t + fx417t shunt system.According to the complainant, the valve was believed to be operating with blockage.The patient underwent a revision procedure.The complainant device was returned to the manufacturer for evaluation.No patient complications were reported as a result of the revision procedure.Patient information: (b)(6) years.Explantation: (b)(6) 2021.The same event: medwatch#3004721439-2021-00133.
 
Additional Manufacturer Narrative
The same event: medwatch#3004721439-2021-00133.Investigation: visual inspection: no abnormalities were detected during the visual inspection.Permeability test: the permeability test showed that the progav 2.0 is permeable.The peritoneal catheter is not permeable.Computer control test: according to our results, we can confirm the presence of an accelerated outflow in the progav 2.0.Adjustability test: the progav 2.0 was found to be adjustable to all pressure settings.Braking force and brake function test: the braking force of the progav 2.0 was within the specified tolerance and the brake function operated as expected.Internal inspection of product: after dismantling of the valve, clearly deposits were found in progav 2.0.Results: based on our investigation results, we can identify an accelerated outflow in the progav 2.0 and a blockage of the peritoneal catheter.We are assuming that the deposits detected have caused the functional impairment.Deposits caused by natural substances in the cerebrospinal fluid, such as protein, blood or tissue particles, are among the known and unavoidable risks and side effects of hydrocephalus therapy.Small amounts of non-visible deposits/proteins might compromise the integrity of the valve.We can exclude a defect at the time of release.The valve met all specifications of the final inspection when released from christoph miethke gmbh & co.Kg.From our point of view, no further regulatory actions are required.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PROGAV 2.0 WITH DISTAL CATHETER
Common Device Name
HYDROCEPHALUS MANAGEMENT
Manufacturer (Section D)
CHRISTOPH MIETHKE GMBH & CO KG
ulanenweg 2
potsdam, 14469
GM  14469
MDR Report Key11849821
Report Number3004721439-2021-00134
Device Sequence Number17091951
Product Code JXG
UDI-Device Identifier04041906136598
UDI-Public4041906136598
Combination Product (Y/N)N
Initial Reporter CountryJA
PMA/510(K) Number
K161853
Number of Events Summarized1
Summary Report (Y/N)N
Device Explanted Year2021
Serviced by Third Party (Y/N)Unknown
Reporter Type Manufacturer
Report Source company representative,user f
Initial Reporter Occupation Other Health Care Professional
Type of Report Initial,Followup
Report Date (Section B) 06/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/02/2025
Device Model NumberFX417T
Device Catalogue NumberFX417T
Device Lot Number20048666
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/17/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/21/2021
Supplement Date Received by Manufacturer06/17/2021
Initial Report FDA Received Date05/19/2021
Supplement Report FDA Received Date06/22/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age59 YR
-
-