| Model Number |
RR 800 |
| Medical Device Problem Code |
Protective Measures Problem (3015)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/30/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
(b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported that the fa quantum controller had an f4 error.It is unknown whether the event happened during surgery and if there was patient involvement.It is unknown if there was a delay or if a backup device was available and how the issue was resolved.No patient injury or other complications were reported.
|
| |
|
Additional Manufacturer Narrative
|
|
Internal complaint reference: (b)(4).H3.H6: the reported device, intended for use in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.A review of the device history records show there are no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.Visual inspection observed a damaged lid.Functional evaluation revealed the unit presents a f4 fault on power up.The unit was opened and found no issues.The complaint was confirmed and the root cause is associated with a component failure.Factors, unrelated to the design or manufacture of the device which could have contributed to the complaint event, include a power surge in the controller or failure of an internal component.A corrective action has been initiated to mitigate future recurrence of similar events.
|
| |
|