The device was not returned for analysis.A review of the lot history record identified no manufacturing nonconformities issued to the reported lot that would have contributed to this event.Additionally, a review of the complaint history did not indicate a lot specific quality issue.It was reported that the procedure was to treat a lesion in the left superficial femoral artery.It should be noted that the graftmaster rapid exchange (rx) coronary stent graft system, domestic, instructions for use (ifu) states: the graftmaster rx is indicated for use in the treatment of free perforations, defined as free contrast extravasation into the pericardium, in native coronary vessels or saphenous vein bypass grafts greater than or equal to 2.75 mm in diameter.It was further reported that the rx graftmaster covered stent was deployed at a maximum pressure of 12 atmospheres (atm).It should be noted in the graftmaster ifu specifies: deploy the stent graft slowly by pressurizing the delivery system in 2-atm increments, every 5 seconds, until the stent graft is completely expanded.Fully expand the stent graft by inflating to nominal pressure at a minimum.The ifu, specifies the nominal pressure is 15 atm.In this case, it is unknown if the ifu deviations directly caused or contributed to the reported event.The investigation was unable to determine a conclusive cause for the reported failure to seal; however, the subsequent treatments appear to be related to the operational context of the procedure.The reported patient effect of dissection, as listed in the graftmaster rx coronary stent graft system ifu, is a known patient effect that may be associated with use of a coronary stent in native coronary arteries.A conclusive cause for the reported patient effect(s) and the relationship to the product, if any, cannot be determined.There is no indication of a product quality issue with respect to the design, manufacture or labeling of the device.The additional graftmaster device referenced is filed under a separate medwatch report number.
|