| Model Number |
QD11-4B |
| Medical Device Problem Codes |
Break (1069); Separation Problem (4043)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
03/23/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
Device was used for treatment, not diagnosis.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.The actual device was returned for evaluation. quality engineering evaluated the device and it was determined that the reported condition was confirmed.The assignable root cause was traced to component failure due to normal wear.A device history review was performed, and no non-conformances were detected related to the reported condition.Udi: (b)(4).
|
| |
|
Event or Problem Description
|
|
It was reported from (b)(6) that during service and evaluation, it was determined that the cutter device was damaged.It was further determined that the cutter device failed visual inspection - showing signs of use and there were broken cutting etches.It was noted in the service order that the cutter device could not be removed from the attachment device.This event did not occur during surgery.There was no patient involvement.There were no reports of injuries, medical intervention or prolonged hospitalization.All available information has been disclosed.If additional information should become available, a supplemental medwatch will be submitted accordingly.
|
| |
|