MAUDE Adverse Event Report: ANGIOMED GMBH & CO. MEDIZINTECHNIK KG VENOVO VENOUS STENT; VENOVO STENT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ANGIOMED GMBH & CO. MEDIZINTECHNIK KG VENOVO VENOUS STENT; VENOVO STENT Back to Search Results
Model Number VENUM20080
Medical Device Problem Code Material Twisted/Bent (2981)
Health Effect - Clinical Code Appropriate Clinical Signs, Symptoms and Conditions Term/Code Not Available (4581)
Date of Event 04/26/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
As the lot number for the device was provided, a review of the device history record is currently being performed.The sample was not returned to the manufacturer for inspection/evaluation.Therefore, the investigation of the reported event is inconclusive.Based upon the available information, the definitive root cause for this event is unknown.The instructions for use (ifu) is adequate for the reported device/patient code(s) and provides general instructions for use, as well as warnings, precautions and potential complications associated with the device.Upon receipt of new or additional information, a follow-up report will be submitted as applicable.(expiration date: 09/2022).
 
Event or Problem Description
It was reported that some time post stent placement, the stent allegedly bent.Reportedly, an additional procedure was required.The current patient status is unknown.
 
Event or Problem Description
It was reported that some time post stent placement the stent allegedly bent.Reportedly, an additional procedure was required.The current patient status is unknown.
 
Additional Manufacturer Narrative
H10: manufacturing review: the lot history record of this lot was reviewed with special attention to the manufacturing and inspection of this product, and the product was found to have met the specification prior to shipment.Investigation summary: the sample was not returned for evaluation.Based on the information available the investigation is closed with inconclusive result.A definitive root cause could not be determined based upon the available information.Labeling review: in reviewing the relevant labeling for this product the potential issue was found addressed.The instructions for use closely describe holding and handling of the system throughout deployment.In regards to pta the instructions for use states: 'predilation of chronic lesions with a balloon dilatation catheter is recommended.', and 'post stent expansion with a balloon dilatation catheter is recommended.' h10: d4 (expiration date: 09/2022).H11: h6 (method).H11: section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text: device not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VENOVO VENOUS STENT
Common Device Name
VENOVO STENT
Manufacturer (Section D)
ANGIOMED GMBH & CO. MEDIZINTECHNIK KG
wachhausstr. 6
karlsruhe 76227
GM  76227
MDR Report Key11853947
Report Number9681442-2021-00323
Device Sequence Number19820690
Product Code QAN
UDI-Device Identifier00801741103858
UDI-Public(01)00801741103858
Combination Product (Y/N)N
Initial Reporter StatePA
Initial Reporter CountryUS
PMA/510(K) Number
P180037
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,consum
Initial Reporter Occupation Non-Healthcare Professional
Type of Report Initial,Followup
Report Date (Section B) 10/21/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Model NumberVENUM20080
Device Catalogue NumberVENUM20080
Device Lot NumberANEW3424
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 04/26/2021
Supplement Date Received by Manufacturer10/21/2021
Initial Report FDA Received Date05/20/2021
Supplement Report FDA Received Date10/23/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
-
-