• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC, INC. PATIENT CONNECTOR; DETECTOR AND ALARM, ARRHYTHMIA

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC, INC. PATIENT CONNECTOR; DETECTOR AND ALARM, ARRHYTHMIA Back to Search Results
Model Number 24967
Device Problem Communication or Transmission Problem (2896)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 04/22/2021
Event Type  malfunction  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported there was intermittent connectivity between the patient connector and the tablet.No testing could be completed on the device, short bursts of continuous communication were enough to program and get the sensing and impedance measurements.Threshold testing could not be completed.The patient connector remains in use.No patient complications have been reported as a result of this event.
 
Manufacturer Narrative
Review of data logs confirmed the customer comment that there was connectivity issues with bluetooth.This complaint was confirmed based on log files.Correction a: patient information correction d: model number, serial number and unique identifier correction d10: concomitant product medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
PATIENT CONNECTOR
Type of Device
DETECTOR AND ALARM, ARRHYTHMIA
Manufacturer (Section D)
MEDTRONIC, INC.
8200 coral sea street ne
mounds view MN 55112
MDR Report Key11854029
MDR Text Key251657371
Report Number2182208-2021-02050
Device Sequence Number1
Product Code DSI
Combination Product (y/n)N
PMA/PMN Number
P890003
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 06/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number24967
Device Catalogue Number24967
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 04/22/2021
Initial Date FDA Received05/20/2021
Supplement Dates Manufacturer Received05/28/2021
Supplement Dates FDA Received06/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
24970A PROGRAMMER; HEAD 24967 QUAL OUS; HEAD 24967 QUAL OUS
Patient Age86 YR
-
-