• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: COVIDIEN LLC XPS SYSTEM NAVIGATED BLADE; BUR, EAR, NOSE AND THROAT

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

COVIDIEN LLC XPS SYSTEM NAVIGATED BLADE; BUR, EAR, NOSE AND THROAT Back to Search Results
Model Number 1884380EM
Medical Device Problem Code Vibration (1674)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/22/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
A healthcare provider (hcp) reported that blade was vibrating more than normal during endoscopic sinus surgery.The surgical tech then re-seated the blade but still had vibrating issue.The procedure was completed with backup product.The blade vibrated before and during use on patient.There was no patient impact.Additional information from rep reported that there were two blades opened with the same issue in which both vibrated before and during use.
 
Additional Manufacturer Narrative
H3: product analysis found that both inner and middle assemblies would rotate freely.The blade was placed into a handpiece with no issues with fitment.The settings were placed on 7,500 rpm in oscillating mode.A concentricity gauge was used on the inner shaft, and the hub was eccentric to the inner shaft.H6: fdm b17, fdr c20 and fdc d14 no longer apply.This is same event as regulatory report #: 9612501-2021-00766.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
XPS SYSTEM NAVIGATED BLADE
Common Device Name
BUR, EAR, NOSE AND THROAT
Manufacturer (Section D)
COVIDIEN LLC
15 hampshire street
mansfield 02048
MDR Report Key11854259
Report Number9612501-2021-00781
Device Sequence Number14414747
Product Code EQJ
UDI-Device Identifier00763000034023
UDI-Public00763000034023
Combination Product (Y/N)N
Initial Reporter StateKS
Initial Reporter CountryUS
PMA/510(K) Number
K130608
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 09/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/09/2023
Device Model Number1884380EM
Device Catalogue Number1884380EM
Device Lot Number0221619316
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer06/03/2021
Type of Report(Section G)Thirty-Day
Initial Date Received by Manufacturer 04/22/2021
Supplement Date Received by Manufacturer08/17/2021
Initial Report FDA Received Date05/20/2021
Supplement Report FDA Received Date09/14/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
-
-