• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: MEDTRONIC IRELAND ATTAIN SELECT II; CATHETER, PERCUTANEOUS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

MEDTRONIC IRELAND ATTAIN SELECT II; CATHETER, PERCUTANEOUS Back to Search Results
Model Number 6248V-130
Medical Device Problem Codes Break (1069); Material Integrity Problem (2978)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 04/28/2021
Type of Reportable Event Malfunction
Additional Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event or Problem Description
It was reported that during the implant procedure when the physician was slitting the inner sheath of the catheter it cracked and fell apart and the physician had to slit the rest of the sheath by hand using a different tool.The catheter was removed.No patient complications have been reported as a result of this event.
 
Additional Manufacturer Narrative
Product event summary: the delivery catheter was returned and analyzed.The analysis indicated that the catheter did not peel along the score lines.There was an issue with the catheter shaft.The mechanical operation of the catheter peeling/slitting/splitting showed a spiral slit.Visual analysis of the delivery catheter indicated damage during use.The analyst noted the delivery catheter was returned without the dilator.Slitting did not start on the score line on the hub of the delivery catheter and did not follow the score line to the distal part of the hub of the delivery catheter.Slitting was restarted at 10.2 cm on the shaft of the delivery catheter.Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTAIN SELECT II
Common Device Name
CATHETER, PERCUTANEOUS
Manufacturer (Section D)
MEDTRONIC IRELAND
parkmore business park west
galway
EI 
MDR Report Key11855209
Report Number9612164-2021-01995
Device Sequence Number3972953
Product Code DQY
UDI-Device Identifier00643169143203
UDI-Public00643169143203
Combination Product (Y/N)N
Initial Reporter StateOR
Initial Reporter CountryUS
PMA/510(K) Number
K123153
Number of Events Summarized1
Summary Report (Y/N)N
Serviced by Third Party (Y/N)N
Reporter Type Manufacturer
Report Source company representative,health
Initial Reporter Occupation Physician
Type of Report Initial,Followup
Report Date (Section B) 06/08/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date02/04/2022
Device Model Number6248V-130
Device Catalogue Number6248V-130
Device Lot Number0010124072
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer05/18/2021
Initial Date Received by Manufacturer 04/28/2021
Supplement Date Received by Manufacturer05/26/2021
Initial Report FDA Received Date05/20/2021
Supplement Report FDA Received Date06/08/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Patient Age84 YR
-
-