| Model Number |
H700123 |
| Medical Device Problem Codes |
Loss of Threshold (1633); Application Program Version or Upgrade Problem (2881)
|
| Health Effect - Clinical Code |
No Clinical Signs, Symptoms or Conditions (4582)
|
| Date of Event |
04/22/2021
|
|
Type of Reportable Event
|
Malfunction
|
|
Additional Manufacturer Narrative
|
|
The investigation revealed that after the workmate claris (b)(4) software upgrade, the workmate claris dws screen turns black/blank and the user needs to reboot the system to regain functionality.
|
| |
|
Event or Problem Description
|
|
During the procedure, there was intermittent pacing and capture.At the start of the case, all 4 pacing channels were selected and assigned to specific poles.Channel 2 was selected first.Pacing behaved as expected.Channel 1 was selected next and pacing appeared to be occurring, but there were only pacing spikes, no capture.Channel 1 was then unassigned and reassigned back to the same pacing site as the original channel 1 site.Pacing began again and the issue was resolved.There was no change in output or catheter location.There were no adverse consequences to the patient.The cause was attributed to a recent upgrade to windows 10 operating software.
|
| |
|