|
Received one (1)139105ext unopened in original packaging.The lot number of the device - 201910044 was verified.A visual inspection found a gap in the seal of the packaging.The packaging was also cut off near the top.Performed a functional inspection, the devices were dye leak tested per (b)(4) which indicated that the packaging had an insufficient heat seal.Dye leaked into the gap in the seal.The manufacturing documents from the device history record have been reviewed with special attention to the manufacturing and inspection of the product.The product released for distribution was found to have met all specifications prior to shipment.This is the only complaint for this lot number and failure mode within the past two years.A two-year review of complaint history revealed there has been a total of 71 complaints, regarding 236 devices, for this device family and failure mode.During this same time frame (b)(4) devices have been manufactured and shipped worldwide.Should all the complaint devices have been found confirmed for this reported failure, the rate of failure would be (b)(4).Per the instructions for use, the user is advised: sterility guaranteed unless package has been opened, broken, or damaged.These devices should never be used when: there is visible evidence of damage to the exterior of the device such as cracked or damaged plastic.This issue will continue to be monitored through the complaint system to assure patient safety.
|
|
During incoming inspection, the distributor rejected this device, 139105ext, for an insufficient heatseal.There was no contact with the patient as this was found during incoming inspection.This will be reported as a malfunction with potential for injury upon reoccurrence.
|