Model Number 72202893
Medical Device Problem Code Difficult to Insert (1316)
Health Effect - Clinical Code No Clinical Signs, Symptoms or Conditions (4582)
Date of Event 05/02/2021
Type of Reportable Event Serious Injury
Additional Manufacturer Narrative
Internal complaint reference case-(b)(4).
 
Event or Problem Description
It was reported that during a cuff suture reconstruction procedure the twinfix ultra were used, due to poor bone quality it was not fully inserted and the introductory cables of the anchors were opened.It was decided to perform an open surgery for the total of the materials.A backup device was available to complete the procedure however, a delay greater than 30 minutes was reported.No patient complications were reported.
 
Additional Manufacturer Narrative
Internal complaint reference: case (b)(4).H3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.A complaint history review concluded this was an isolated event.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A review of risk management files found that the reported failure was documented appropriately.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX ULTRA TI 4.5 W/2 UB (WH & BLUE)
Common Device Name
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key11856984
Report Number1219602-2021-01181
Device Sequence Number17309070
Product Code MBI
UDI-Device Identifier03596010652287
UDI-Public03596010652287
Combination Product (Y/N)N
Initial Reporter CountryBR
PMA/510(K) Number
K100159
Number of Events Summarized1
Summary Report (Y/N)N
Reporter Type Manufacturer
Report Source company representative,foreig
Initial Reporter Occupation Other
Type of Report Initial,Followup
Report Date (Section B) 06/28/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Operator of Device Health Professional
Device Expiration Date07/29/2025
Device Model Number72202893
Device Catalogue Number72202893
Device Lot Number2058878
Was Device Available for Evaluation? No
Initial Date Received by Manufacturer 05/03/2021
Supplement Date Received by Manufacturer06/25/2021
Initial Report FDA Received Date05/20/2021
Supplement Report FDA Received Date06/28/2021
Is This a Single-Use Device that was
Reprocessed and Reused on a Patient? (Y/N)
No
Patient Sequence Number1
Outcome Attributed to Adverse Event Required Intervention;
Patient Age39 YR
-
-